II-41
d. Provisional measurement check in single range mode
Display Menu No.
: CHK-3 : RANGE
This function is used to set the output range (+/-. Hi/Low).
When provisional range setting is activated, mark < R > appears at the far right of the 1
st
line,
except when [STATUS 1] or [STATUS 2] is displayed. After executing [0: FINISH], the < R >
mark disappears.
Note:
- If power to the flowmeter is shut down while this function is activated, this parameter will remain
active when the system is rebooted.
e. Checking single path measurement while multi-path measurement is activated
Display Menu No.
: CHK-4 : MULTI PATH
Select an active measurement path (1
st
path, 2
nd
path, 3
rd
path, or 4
th
path).
Mark < M > appears at the far right of the 1
st
line of the display when activating a measurement
path except when [STATUS 1] or [STATUS 2] is displayed. The < M > mark disappears after [0:
FINISH] is executed.
Note:
- If power to the flowmeter is shut down while this function is activated, this parameter will remain
active when the system is rebooted.
-
While this function is activated, the SMOOTHING parameter, “INP-16: SMOOTHING”
related to measuring path(s) is determined as follows.
[Input time (0 to 120sec)]/[Total number of measuring paths (1 to 4) input]
Содержание UFL-30
Страница 1: ...Doc No KF08 001J Ultrasonic Flowmeter UFL 30 Installation Operation Manual ...
Страница 11: ... 10 blank ...
Страница 12: ...I Installation ...
Страница 60: ...II Operation ...
Страница 109: ...II 50 blank ...
Страница 110: ...III Other ...
Страница 123: ...III 14 III 2 5 Dimensions Main Unit ...
Страница 128: ...III 19 blank ...
Страница 129: ...III 20 blank ...
Страница 145: ...III 36 blank ...
Страница 165: ...III 56 blank ...