II-34
(4) Integration (Totalizing) Function
a. Changing totalizing unit
Display Menu No.
: INP-211 : +ITG.UNIT
INP-212 : -ITG.UNIT
Initial value
: “0 : NONE” (no integration)
For mass units (i.e. kg, t, etc.), the mass flow rate unit (i.e. kg/h, t/min., etc.) must be selected
prior to selecting the totalizing unit under parameter, INP-122: FLOW UNIT.
b.
Totalizing start and end
Display Menu No.
: INP-221 : STOP MODE
Initial value
: “0 : MANUAL” (Continuous Integration)
Display Menu No.
Function
INP-222 : [START]
Integration Start
INP-223 : [PRESET START]
Integrating start from input value (refer to (4)-c)
INP-224 : [STOP]
Integration Stop
INP-225 : TIMER
Integrating during input time, automatic stop.
Range of values: 00h00m00s to 99h59m059s.
When the totalizing function is active, the reverse field of the “I” mark blinks at the far right of the
1
st
line of the display, except during [STATUS 1] or [STATUS 2] display. The “I” mark disappears
when the totalizing function stops.
Note: The blinking cursor shifts through an hours, minutes,
and seconds (“h”, “m”, and “s”) time sequence.
0
1
h
2
3
m 4
5
s
Individually set up for both forward
and backward modes
Содержание UFL-30
Страница 1: ...Doc No KF08 001J Ultrasonic Flowmeter UFL 30 Installation Operation Manual ...
Страница 11: ... 10 blank ...
Страница 12: ...I Installation ...
Страница 60: ...II Operation ...
Страница 109: ...II 50 blank ...
Страница 110: ...III Other ...
Страница 123: ...III 14 III 2 5 Dimensions Main Unit ...
Страница 128: ...III 19 blank ...
Страница 129: ...III 20 blank ...
Страница 145: ...III 36 blank ...
Страница 165: ...III 56 blank ...