II-27
d. Prevent frequent changeover between Forward measuring and Backward measuring
during low flow situation
d-1 Forward / Backward Operation
Display Menu No. : INP-146 : FW/BW OPE.
Initial value
: “ 0: OFF ”
d-2 Low Flow Cut
Display Menu No. : INP-153 : +LOW CUT
: INP-156 : -LOW CUT
Initial value
: 1.5% of Full Scale or User defined value
Range of Input
: +LOW CUT
: 0.000 to 9999.00
: -LOW CUT
: -9999.0 to 0.000
Flow Rate Unit
: According to Measuring Unit that was set in
INP-121 : MULTIPLIER &INP-122 : FLOW UNIT.
Note :
In case of input into “-LOW CUT”, the character at the left-end in the LCD can be a “blank” or
“minus”.
Individually setup
both Forward and Backward
Содержание UFL-30
Страница 1: ...Doc No KF08 001J Ultrasonic Flowmeter UFL 30 Installation Operation Manual ...
Страница 11: ... 10 blank ...
Страница 12: ...I Installation ...
Страница 60: ...II Operation ...
Страница 109: ...II 50 blank ...
Страница 110: ...III Other ...
Страница 123: ...III 14 III 2 5 Dimensions Main Unit ...
Страница 128: ...III 19 blank ...
Страница 129: ...III 20 blank ...
Страница 145: ...III 36 blank ...
Страница 165: ...III 56 blank ...