III-27
(5) Specification for serial communication
5-1. Hardware settings
DIGITAL port1 of main unit communication specifications are provided in Table 5.1.
Table 5.1
Communication Parameters
Value
Standard
EIA-232
(
RS-232C
)
Baud rate
note1
4800bps, 9600bps, 19200bps
( Default; 9600bps )
Data bit
8
Parity Even
Stop bit
1
Flow control
None
Output interval
note2
Every 0
~
3600 seconds
No output in case of "0" setting.
(Default 1sec interval)
Note1
Baud rate for port 1 is set under MENU [ INP-421 : BAUD RATE#1 ] by the main unit LCD
key-pad only.
Select Menu INP-421, then enter baud rate selection as below.
3 speeds are selectable.
0: 4800bps
1: 9600bps (Default)
2: 19600bps
Note2
Output interval is set under MENU [ INP-41 ] by the main unit LCD key-pad or through
commissioning software at [ Advanced - INPUT - Digital Output ].
Select Menu INP-41 from LCD Key-pad For software, set under [INPUT-Digital output].
Enter INP-41, data settings viewable.
Manually change settings.
I N P - 4
2
< |
( B A U D
R A T E )
I N P - 4 2
1
| >
B A U D
R A T E # 1
B A U D
R A T E # 1
1 :
9 6 0 0 B P S
I N P - 4
1
| >
I N T E R V A L
I N T E R V A L
0 0 0 1
s e c
I N T E R V A L
0
0 0 1
s e c
Содержание UFL-30
Страница 1: ...Doc No KF08 001J Ultrasonic Flowmeter UFL 30 Installation Operation Manual ...
Страница 11: ... 10 blank ...
Страница 12: ...I Installation ...
Страница 60: ...II Operation ...
Страница 109: ...II 50 blank ...
Страница 110: ...III Other ...
Страница 123: ...III 14 III 2 5 Dimensions Main Unit ...
Страница 128: ...III 19 blank ...
Страница 129: ...III 20 blank ...
Страница 145: ...III 36 blank ...
Страница 165: ...III 56 blank ...