I-13
Fig 1-2-4-2. Wire Entry & Earth connector
Fig 1-2-4-3. Internal Connection Terminal of Main Unit (AC Power)
NOTE
In case of DC power supply,
ACF-30
PCB will be changed to
DCF-30
PCB.
Connection will be as below.
AC(L) to be +DC
AC(N) to be -DC
PE to FE
NOTE
Terminal TB1 and TB3-9 operation must be handled by screwdriver type slotted 3mm~3.5mm as
tool for lever operation.
These terminals covers 5~6mm wire length as peeled off.
Functional earth terminal
(
Grounding requires
Type D or higher
)
Housing bottom
Cable entry for power cable
Cable entry for
transducer cable
for CH1
Cable entry for
transducer cable
for CH2
or I/O cables
Cable entry for I/O cable