I-20
(2) 4-Path Connection
(If ordered as 4-path system, skip to 2-E)
2-A. Remove the shield cover of transducer cable terminal
Remove the shield cover (indicated below) for multi-path set up (bottom diagram) by loosening
the screws.
In case of ordered as 4-path system originally, shield cover and clamp stand are not included.
2-B. Pulser Module
With 1 or 2 modules already in slot CH1 (and CH2),
extra pulser modules are required for 4-path
measurement. (CH1 to CH4)
These modules should be set on the mother board next
to the fourth slot .
2-C. Module Insertion
In order to insert the extra modules, the module cover
must first be removed by loosening the screws.
Insert modules, paying attention to the orientation of the
board.
CAUTION
Pulser Module may be defected by static electricity of your hand. Discharge it prior to touch the module.
2-D. Replace module cover
Replace module cover after insertion of the extra modules.
2-E. Check Terminals for transducer line
One "multi-path junction box" is required for 4-path measurement.
Connection should be made to the terminals for transducer connectors
as shown in Photo 2 through the inlet grommet. Final installation image is as below.
Remove
Shield Cover
Photo for 2-B
Terminals for Transducer
connector
1U1D
2U2D 3U3D 4U4D
Terminals for transducer
C
H
1
CH2~4
Содержание UFL-30
Страница 1: ...Doc No KF08 001J Ultrasonic Flowmeter UFL 30 Installation Operation Manual ...
Страница 11: ... 10 blank ...
Страница 12: ...I Installation ...
Страница 60: ...II Operation ...
Страница 109: ...II 50 blank ...
Страница 110: ...III Other ...
Страница 123: ...III 14 III 2 5 Dimensions Main Unit ...
Страница 128: ...III 19 blank ...
Страница 129: ...III 20 blank ...
Страница 145: ...III 36 blank ...
Страница 165: ...III 56 blank ...