II-2-8 Error Message
If invalid data or selection is entered, an error message will be displayed. The previous setting is
automatically retained and the display will revert to the previous indication.
ex : When invalid data is input
S M O O T H I N G
0
1 5
s e c
S M O O T H I N G
1 5
0
s e c
S M O O T H I N G
I N P U T
E R R O R
Range of Input : 0 to 120sec
ex : When an invalid choice is selected
I M P - 1 3
3
< >
+ L O W
V A L U E
( x )
I M P - 1 3 3
< >
I N V A L I D
( x )
ex : When measured flow rate is over the maximum flow rate as set by + MAX VALUE parameter
(INP-132 in A-OUT) or when converted velocity e/-30m/s.
Provided that +MAX VALUE (INP-132) was 100.00,
V o l .
< A R M > B R D O
I
1 0 0 . 0 0
m
3
/ h
V o l .
< A R M > B R D O
I
F . S .
m
3
/ h
* According to each setting
Over +MAX VALUE (more than 100.01)
“F.S.” = Full Scale = Analog Output Range
ex : When values are over digits
Supposing that iMAX VALUE (INP-132) = 100 and
initial -MAX VALUE (INP-134) = -100. If +MAX VALUE is
changed to 0.300, according to the decimal places of +MAX
VALUE, a -MAX VALUE of -100.000 should be indicated.
However the maximum no. of digits of the LCD for this string is 7, including decimal point and +/-
sign. Thus -100.000, at 8 digits, will be deemed “Over Digits” and the -MAX VALUE display
indication will appear as [ ------- ].
ENT
Mark (X) indicates
invalid choice.
LCD indicates
100.00 and F.S.
alternately.
ENT
Содержание UFL-30
Страница 1: ...Doc No KF08 001J Ultrasonic Flowmeter UFL 30 Installation Operation Manual ...
Страница 11: ... 10 blank ...
Страница 12: ...I Installation ...
Страница 60: ...II Operation ...
Страница 109: ...II 50 blank ...
Страница 110: ...III Other ...
Страница 123: ...III 14 III 2 5 Dimensions Main Unit ...
Страница 128: ...III 19 blank ...
Страница 129: ...III 20 blank ...
Страница 145: ...III 36 blank ...
Страница 165: ...III 56 blank ...