I-24
I-2-6 Functional ground connection
System should be grounded as near to the main unit as possible based on the grounding
specifications in (*1). Refer to Fig.1-2-4-2 for functional ground connection points.
Functional grounding would be worked for noise shield and lightning surge protection.
(*1) Grounding
Grounding resister: Less than 100 ohm
Applicable grounding material: Steel Frame & Metal Exterior for low voltage
(less than 300V)
Thickness of grounding cable: Diameter more than 1.6mm
2
(Soft-copper wire)
I-2-7 Installation of disconnection device for power line
DANGER
Be sure to stop power supply while installation of the device to prevent electrical shock .
This flowmeter does not have a disconnection device which meets the relevant requirements of IEC
60947-1 and IEC 60947-3 for power line. Therefore, install a switch or circuit-breaker as the means for
disconnection of power line for AC power type of main unit to comply with EC directive.
It is specified as follows,
a) a switch or circuit-breaker shall be included in the building installation;
b) it shall be in close proximity to the equipment and within easy reach of the OPERATOR;
c) it shall be marked as the disconnecting device for the equipment.
NOTE
Required ratings for the flowmeter are 2A rated current and 250V rated voltage.
I-2-8 Insulation for DC power source
The DC power source must be separated from mains by means of reinforced insulation for DC power
type of main unit to comply with EC directive.
Содержание UFL-30
Страница 1: ...Doc No KF08 001J Ultrasonic Flowmeter UFL 30 Installation Operation Manual ...
Страница 11: ... 10 blank ...
Страница 12: ...I Installation ...
Страница 60: ...II Operation ...
Страница 109: ...II 50 blank ...
Страница 110: ...III Other ...
Страница 123: ...III 14 III 2 5 Dimensions Main Unit ...
Страница 128: ...III 19 blank ...
Страница 129: ...III 20 blank ...
Страница 145: ...III 36 blank ...
Страница 165: ...III 56 blank ...