I-15
(4) Connect transducer cables
Connect the coaxial cable for the upstream transducer to a Terminal for transducer. Likewise
connect the cable for the downstream transducer to a same terminal.
For Multi-path measurement connection, please refer to I-2-5.
For the transducer cable treatment, refer to clause I-2-10.
aa. Remove the sensor terminal cover and the cable clamps for ch1-UP and ch1-DOWN of Path-1.
bb. Loose the screw on both sides of the sensor cable terminal "TB10" by a screwdriver.
(Two screws at center should not be used. They are connected to the signal ground.)
Pass the "UP" sensor coaxial cable through the cable gland of upper left.
cc. Put GND part of the cable on the clamp stand and pass the core conductor of the coaxial cable
under the board spring of the screw.
Signal for UP
Signal for DN
GND
Terminal cover
Cable clamps
clamp
stand