I-23
Step3: Fix cable gland on main unit.
Step4: Fix cable gland on junction box
as shown below.
side as shown below.
2-K. Connection of Transducer lines from junction box to terminals of main unit
Connect the 8 transducer lines to the main unit. Each line is color-coded. Connect lines to their
proper terminals.
1U = 1
st
path upstream side
1D = 1
st
path downstream side
2U = 2
nd
path upstream side
2D = 2
nd
path downstream side
3U = 3
rd
path upstream side
3D = 3
rd
path downstream side
4U = 4
th
path upstream side
4D = 4
th
path downstream side
2-L. Transducer line connection to junction box
After installing junction box near to the main unit, connect transducer lines to the junction box.
CAUTION
After installation and wiring work, tighten the cable glands exactly to protect against water and dust.
1U
1D
2U
2D
3U
3D
4U
4D