I-17
(5) Connect I/O cable (analog or relay output) line
Use high quality standard issue cables with a cross-section of 0.75 - 2 mm
2
as signal cables which
outer size of cable is 6~12mm as completed treatment.
.
aa. Pass the I/O cable through the cable gland of upper second from right. After positioning the
cable, tighten the cable gland exactly to fix the cable and to protect against water and dust .
(Torque : Approx. 1.5 N m)
bb. Insert each wire in the loading slot while pushing the lever of the terminal TB6 and TB8 by the
screwdriver, type slotted.
Sample of connection Analog output CH1 and Relay output CH4 are shown below.
Analog output (A-OUT)
Relay output
Finish
Positive-line ----> #1 slot of TB8
Positive-line -----> #1 slot of TB6
Negative-line ---> #2 slot of TB8
Negative-line ----> #2 slot of TB6
CAUTION
Check conformity for the polarity of terminal and wire
and for rating of connected device. It may be that wrong
wiring or wrong rating will break the equipment.