II-30
b. Change digits of flow rate display and/or measuring range
Display Menu No. : INP-132 : +MAX VALUE
The number of maximum characters on the LCD and the position of the decimal point will be
arranged according to the parameter, Maximum Flow Rate (Forward).
The maximum number of characters on LCD is 7 characters including “ - “ mark & a decimal
point.
(Provided that maximum range of flow rate is 99999.0 regardless of measuring unit.)
Each function of High/Low Alarm, Zero Shift and Low Flow Cut will be also in accordance with
same condition.
Note :
It can be set to four decimal places. The fifth decimal place is not available.
Analog Output Pattern
Display Menu No.
for Measuring Range
Range of Value
Single Range
INP-132 : +MAX VALUE
0.001 to 99999.0
INP-134 : -MAX VALUE
-99999 to -0.001
Double Range
High Range
INP-132 : +MAX VALUE
0.001 to 99999.0
INP-134 : -MAX VALUE
-99999 to -0.001
Low Range
INP-133 : +LOW VALUE
0.001 to 99999.0
INP-135 : -LOW VALUE
-99999 to -0.001
Flow Rate Unit
According to Measuring Unit that was set in INP-121 : MULTIPLIER &INP-122 : FLOW UNIT.
Note :
The decimal place is according to the parameter of INP-132 : +MAX VALUE.
The character at the left-end in the LCD can be a “blank” or “minus”.
Содержание UFL-30
Страница 1: ...Doc No KF08 001J Ultrasonic Flowmeter UFL 30 Installation Operation Manual ...
Страница 11: ... 10 blank ...
Страница 12: ...I Installation ...
Страница 60: ...II Operation ...
Страница 109: ...II 50 blank ...
Страница 110: ...III Other ...
Страница 123: ...III 14 III 2 5 Dimensions Main Unit ...
Страница 128: ...III 19 blank ...
Страница 129: ...III 20 blank ...
Страница 145: ...III 36 blank ...
Страница 165: ...III 56 blank ...