III-30
III-2-8 Digital output profiles (MODBUS Protocol)
(1) Overview
Flowmeter can provide digital data output of flowrate, velocity data, totalizing value as well as
measurement status.
This digital data is output from DIGITAL port1 via EIA-232 (RS-232C) connection by MODBUS
protocol when selected by commissioning software.
(2) Wiring Connection
Physical wiring between Main unit and digital output is as follows,
Main unit setting is done through Digital Port 2 by intercommunication mode under TOKYO KEIKI
original protocol.
MODBUS communication is output from Digital Port 1 in accordance with main unit setting.
CAUTION
1. Connection cable length must be less than 3 m long to comply with EC directive.
2. Do not connect the connection cable to LAN connector of the personal computer, that may
damage the flowmeter main unit and/or the personal computer.
3. Make sure that connections are made correctly as incorrect connections may result in
damage to the main unit and connected accessories.
NOTE; Each “RTS and CTS”, “DTR and DTS” of Digital Port 1 is shorted on flowmeter ‘s main unit side.
RJ-45 side
1 Blue
2.Orange
3 Black
4 Red
5 Green
6 Yellow
7 Brown
8 White
1 DCD
2.RxD
3 TxD
4 DTR
5 GND
6 DSR
7 RTS
8 CTS
D-sub 9 pin side
1 Blue
2.Orange
3 Black
4 Red
5 Green
6 Yellow
7 Brown
8 White
1
2
3
4
5
6
7
8
RJ-45 side
D-sub 9 pin side
PC with
Commissioning software
RJ-45 / RS-232C
Converter
MODBUS Output
or
D-OUT Output
RJ-45 Cable
(Straight Cable)
Содержание UFL-30
Страница 1: ...Doc No KF08 001J Ultrasonic Flowmeter UFL 30 Installation Operation Manual ...
Страница 11: ... 10 blank ...
Страница 12: ...I Installation ...
Страница 60: ...II Operation ...
Страница 109: ...II 50 blank ...
Страница 110: ...III Other ...
Страница 123: ...III 14 III 2 5 Dimensions Main Unit ...
Страница 128: ...III 19 blank ...
Страница 129: ...III 20 blank ...
Страница 145: ...III 36 blank ...
Страница 165: ...III 56 blank ...