III-7
Digital output St’d/option
Standard
Output 1
One-way output mode
Following data is output per set output cycle
Instantaneous flow rate, forward/backward flow totalized value
and various warnings.
(flow meter mode: linear flow rate and various
warnings)
Note: Instantaneous flow rate will change to velocity value in case of velocity
mode. No totalized values available.
Output 2
Intercommunication mode
Connection to PC enables setting of flowmeter unit, setting menu
and reading of measurement values and operation status.
Output type
RS232C (non-insulated output
Output cycle
1 to 3600 seconds possible. (output 1 only)
Communication
speed
4800 bps, 9600 bps or 19200 bps selectable
Data bit length
8 bit/1 stop bit
Parity check
EVEN
Format
TOKYO KEIKI Original Format (ASCII)
MODBUS (Selectable; only for Output1)
Synchronization Asynchronous
Cable length
Up to 3m
Note: To comply with EC directives, use less than 3m cable.
Multi-path
measurement
St’d/option
Option
Quantity
2 path or 4 path
Details
2 path:
- 1 additional pulser module required and installed in main unit
(total 2 modules).
- Transducer, fixture, extension cable for each path required.
4 path:
- External multi-path junction box is required and 2 special
coaxial composite cables are connected to the flowmeter main
unit.
- Transducer cable connected in junction box by BNC
connector.
- 3 additional pulser modules required and installed in main unit.
(total 4 modules).
- Transducer, fixture, extension cable for each path required.
Содержание UFL-30
Страница 1: ...Doc No KF08 001J Ultrasonic Flowmeter UFL 30 Installation Operation Manual ...
Страница 11: ... 10 blank ...
Страница 12: ...I Installation ...
Страница 60: ...II Operation ...
Страница 109: ...II 50 blank ...
Страница 110: ...III Other ...
Страница 123: ...III 14 III 2 5 Dimensions Main Unit ...
Страница 128: ...III 19 blank ...
Страница 129: ...III 20 blank ...
Страница 145: ...III 36 blank ...
Страница 165: ...III 56 blank ...