II-8
2. 1:+/-:SNGL (1 system, 2 direction single range output)
Measures forward and reverse direction flows. Measurement ranges for each direction can be
independently set.
Graph of flow relative to output current is shown below.
3. 1:+:DBL (1 system, 1 direction double range output)
Measures forward direction flow only. Same current values are output from channels 1 and 2.
Graphs of flow relative to output current are shown below.
Automatic double range
4. 1:+/-:DBL (1 system, 2 channel double range output)
Measures flows both in forward and reverse directions.
Measurement ranges for each direction can be independently set..
Graphs of flow relative to output current are shown below.
Automatic double range