I-16
dd. Hold down the GND part of the coaxial cable by the cable clamp and tighten up the screw firmly.
After that, tighten up the screw of the terminal "TB10" without so strong.
Note that this turn of work is important. When the wrong order, work to clamp GND maybe give
a stress to the core conductor of the cable and cut it.
ee. The "DOWN" sensor coaxial cable is similarly processed from aa to dd.
ff. Close the terminal cover.
gg. Finally, tighten the cable gland exactly to fix the cable and to protect against water and dust.
(Torque : Approx. 1.5 N m)