Synergy Controller Technical Manual, Revision H
Page 59
L11 Dehumidify / Vent On
Percent dehumidify required to enable dehumidify device. (0 to 100%)
L12 Dehumidify / Vent Off
Percent humidify required to turn off dehumidify device. (0 to 100%)
L14 Time Delay Boost Cool
Time delay required before Boost Cool is enabled. (0 to 1200 seconds)
L15 Compressor Turn-Off Delay
Delay required before turning off a compressor. (0 to 5 minutes)
LEV1 Drier / Dehumidify Coil
Selects Drier or the Dehumidify Coil for dehumidification. (Dehumidify Coil, Drier). When equipped
with a drier, the chamber will typically be capable of achieving a lower humidity.
Содержание Synergy Compact
Страница 14: ...Synergy Controller Technical Manual Revision H Page 14 2 0 SPECIFICATIONS 2 1 Data Sheet...
Страница 71: ...Synergy Controller Technical Manual Revision H Page 71...
Страница 106: ...Synergy Controller Technical Manual Revision H Page 106...
Страница 123: ...Synergy Controller Technical Manual Revision H Page 123...
Страница 141: ...Synergy Controller Technical Manual Revision H Page 141 TE1151 6 Triac Output Board Schematic...
Страница 142: ...Synergy Controller Technical Manual Revision H Page 142 TE1708 6 Relay Board Schematic...
Страница 188: ...Synergy Controller Technical Manual Revision H Page 188...
Страница 274: ...Synergy Controller Technical Manual Revision H Page 274 Synergy Controller Step Syntax...
Страница 294: ...Synergy Controller Technical Manual Revision H Page 294...
Страница 312: ...Synergy Controller Technical Manual Revision H Page 312...
Страница 345: ...Synergy Controller Technical Manual Revision H Page 345...
Страница 346: ...Synergy Controller Technical Manual Revision H Page 346...