Synergy Controller Technical Manual, Revision H
Page 161
Open the SETUP\Logging\Data\ Folder and
select Channel Readings.
Enable Channel 1 Actual and Channel 2
Actual (Select the item and press the
"Change" Button)
Channel 1 Setpoint
Enable
Channel 2 Setpoint
as required
Channel 3 Setpoint
as required
Note that this folder may have up to four
Setpoints listed.
PID Output
Enable PID Output Logging for
CH1, Cool and Heat folder.
These constants can optionally be logged
Proportional Band
a PID Constant
Reset
This is also a PID Constant
Once the logging settings are adjusted, run
the chamber with a representative profile
that includes the test conditions required for
the application.
Note that PID variables
Cycle Time, Pn, In,
Dn
and
Error
can also be logged but are not
required logged for tuning applications
Содержание Synergy Compact
Страница 14: ...Synergy Controller Technical Manual Revision H Page 14 2 0 SPECIFICATIONS 2 1 Data Sheet...
Страница 71: ...Synergy Controller Technical Manual Revision H Page 71...
Страница 106: ...Synergy Controller Technical Manual Revision H Page 106...
Страница 123: ...Synergy Controller Technical Manual Revision H Page 123...
Страница 141: ...Synergy Controller Technical Manual Revision H Page 141 TE1151 6 Triac Output Board Schematic...
Страница 142: ...Synergy Controller Technical Manual Revision H Page 142 TE1708 6 Relay Board Schematic...
Страница 188: ...Synergy Controller Technical Manual Revision H Page 188...
Страница 274: ...Synergy Controller Technical Manual Revision H Page 274 Synergy Controller Step Syntax...
Страница 294: ...Synergy Controller Technical Manual Revision H Page 294...
Страница 312: ...Synergy Controller Technical Manual Revision H Page 312...
Страница 345: ...Synergy Controller Technical Manual Revision H Page 345...
Страница 346: ...Synergy Controller Technical Manual Revision H Page 346...