The liquid level sensor is unable to detect foam. In case of foaming:
1.
Empty the waste bottle as soon as the foam level has reached the
maximum filling level indicated on the waste bottle.
2.
Add a commercially available anti-foaming agent, such as silicone
oil, to the empty waste bottle to reduce foaming. Use concentrations
of anti-foaming agents as recommended by the manufacturers.
3.
Consider switching to a larger waste bottle and additionally
increasing the concentration of anti-foaming agent in the waste
bottle.
4.
Carefully swirl the waste bottle from time to time to improve
mixing between the foam layer and anti-foaming agent.
To connect the power supply cable:
Warning
Do not operate your instrument from a power outlet that has
no ground connection. Do not use a power supply cable other than the
Thermo Scientific power supply cable designed for your region.
▲
1.
Ensure that the mains (
I
/
O
) switch (Figure 2–21) at the right side
panel of the instrument is in the off (
O
) position.
2.
Connect the power supply cable to the power supply connector.
3.
Connect the power supply to a correctly installed line power outlet
with a grounded conductor.
Foaming
Connecting the
power supply cable
Содержание Wellwash Versa
Страница 1: ...Thermo Scientific Wellwash Versa User Manual Rev 1 3 ...
Страница 2: ......
Страница 3: ...Thermo Scientific Wellwash Versa User Manual Rev 1 3 Cat No N11166 ...
Страница 40: ...Wellwash Versa Main Parts Shaker 38 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 130: ...Maintenance Disposal of the instrument 128 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 142: ...System Log 140 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 144: ...Certificate of Decontamination 142 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 146: ......
Страница 147: ...N11166 Thermo Fisher Scientific Oy Ratastie 2 P O Box 100 FI 01621 Vantaa Finland www thermoscientific com ...