Maintenance
Cleaning the plate carrier
116
Thermo Scientific Wellwash Versa User Manual
Thermo Fisher Scientific
To remove the priming vessel:
1.
Push the priming vessel away from yourself towards the instrument
until you hear a snap.
2.
Lift the priming vessel up and remove it (Figure 11–42
).
Figure 11–42.
Removing the priming vessel
3.
Clean the priming vessel with a mild detergent and rinse with
deionized distilled water.
4.
Replace the priming vessel by inserting it downwards and pulling it
towards yourself.
Warning
The priming vessel may be contaminated after the instrument
has been used.
▲
To clean the plate carrier:
Caution
Do not disassemble the plate carrier.
▲
1.
Switch the power off and unplug the instrument.
2.
Use disposable powder-free gloves.
3.
If spillages have occurred, move the plate carrier by pushing it to
the left (Figure 11–43). This makes it easier for you to clean the
area under the plate carrier.
Cleaning the
plate carrier
Содержание Wellwash Versa
Страница 1: ...Thermo Scientific Wellwash Versa User Manual Rev 1 3 ...
Страница 2: ......
Страница 3: ...Thermo Scientific Wellwash Versa User Manual Rev 1 3 Cat No N11166 ...
Страница 40: ...Wellwash Versa Main Parts Shaker 38 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 130: ...Maintenance Disposal of the instrument 128 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 142: ...System Log 140 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 144: ...Certificate of Decontamination 142 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 146: ......
Страница 147: ...N11166 Thermo Fisher Scientific Oy Ratastie 2 P O Box 100 FI 01621 Vantaa Finland www thermoscientific com ...