Installation
Setups
Thermo Fisher Scientific
Thermo Scientific Wellwash Versa User Manual
23
Caution
Do not cover the venting holes.
▲
Caution
If foaming occurs in the waste bottle, refer to “Foaming” on
page 27.
▲
Caution
Do not limit, change or remove the float position on the liquid
level sensor.
▲
Figure 2–16.
Correct position of the float on the buffer bottle sensor
Smaller bottles can also be used with an optional bottle holder
(Cat. No. N10820).
Caution
If a smaller size buffer bottle is used with the bottle stand, the
remaining volume in the bottle after the liquid level sensor indicates
empty may be smaller than the nominal 290 ml.
▲
Caution
Before disconnecting the tubing, prime the liquid system with
air to avoid liquid spillage.
▲
Figure 2–17 shows the connections of the liquid level sensor cabling and
liquid bottle tubing.
Connection diagram
Float
Содержание Wellwash Versa
Страница 1: ...Thermo Scientific Wellwash Versa User Manual Rev 1 3 ...
Страница 2: ......
Страница 3: ...Thermo Scientific Wellwash Versa User Manual Rev 1 3 Cat No N11166 ...
Страница 40: ...Wellwash Versa Main Parts Shaker 38 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 130: ...Maintenance Disposal of the instrument 128 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 142: ...System Log 140 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 144: ...Certificate of Decontamination 142 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 146: ......
Страница 147: ...N11166 Thermo Fisher Scientific Oy Ratastie 2 P O Box 100 FI 01621 Vantaa Finland www thermoscientific com ...