Maintenance
Cleaning the wash head(s)
Thermo Fisher Scientific
Thermo Scientific Wellwash Versa User Manual
119
To clean the wash head(s), refer to features in the
Maintenance
menu:
“Clean wash head” on page 90, “Soak wash head” on page 92, and/or
“Move wash head” on page 93.
Wash head tips that are clogged can be cleaned using a declogging tool
(Figure 11–45). The
thin declogging tool
is intended for declogging a
2x8 cell wash head and/or a 1x16 wash head and the
normal declogging
tool
for declogging other wash heads than the above-mentioned.
Warning
The wash heads may be contaminated after the instrument
has been used.
▲
Warning
Disinfect the wash heads thoroughly before you remove
them.
▲
Warning
Prime the instrument without liquid (with air) to empty all
liquid from the system before removing the wash heads.
▲
Warning
When handling the waste bottle, it is advisable to observe
applicable safety precautions, such as the wearing of disposable powder-
free gloves, safety glasses, and protective clothing, to avoid potential
infectious disease contamination.
▲
Caution
Clean the wash head tips regularly. Wash head tips that are
clogged can be cleaned using a declogging tool (Figure 11–45).
▲
Figure 11–45.
Thin and normal declogging tools
Check that the wash head tips are in good condition and not deformed.
●
Detach the wash head from the tubing.
●
Remove the wash head plugs and use the declogging tool to clean
the tips.
Cleaning the
wash head(s)
Mechanical
cleaning of the
wash head
Содержание Wellwash Versa
Страница 1: ...Thermo Scientific Wellwash Versa User Manual Rev 1 3 ...
Страница 2: ......
Страница 3: ...Thermo Scientific Wellwash Versa User Manual Rev 1 3 Cat No N11166 ...
Страница 40: ...Wellwash Versa Main Parts Shaker 38 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 130: ...Maintenance Disposal of the instrument 128 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 142: ...System Log 140 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 144: ...Certificate of Decontamination 142 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 146: ......
Страница 147: ...N11166 Thermo Fisher Scientific Oy Ratastie 2 P O Box 100 FI 01621 Vantaa Finland www thermoscientific com ...