Installation
Installing the aerosol cover
Thermo Fisher Scientific
Thermo Scientific Wellwash Versa User Manual
31
The transparent aerosol cover may be present (Figure 3–24) or absent
during operation. The aerosol cover primarily protects the user against
biohazardous aerosols and the site against environmental contamination.
Figure 2–23 shows the aerosol cover being installed. The corners have
magnets for quick magnetic mounting of the cover.
Figure 2–23.
Installing the aerosol cover
Switch the instrument on. The system performs initialization tests (=
self diagnostics) each time it is switched on.
To change the preferred user interface language or the local date/time,
refer to “Settings menu” on page 81.
Caution
Do not switch the power off nor plug/unplug the USB
memory device during “Performing self diagnostics”.
▲
Caution
Do not touch the wash head or plate carrier when the
instrument is in use.
▲
Note
Ensure that the wash heads are installed.
▲
Before putting the instrument into use, perform the following
operational check. For more information, refer to Chapter 4: “
Operating
the Instrument
”.
●
Ensure that the liquid bottles are empty.
●
Turn the empty bottles upside down one by one to ensure that the
liquid level sensor in the bottle works properly. The liquid level icon
should be
green
when the buffer bottle is upside down and
red
when the buffer bottle is in an upright position. The liquid level
Installing the
aerosol cover
Switching on
Performing the
operational
check after
switching on
Содержание Wellwash Versa
Страница 1: ...Thermo Scientific Wellwash Versa User Manual Rev 1 3 ...
Страница 2: ......
Страница 3: ...Thermo Scientific Wellwash Versa User Manual Rev 1 3 Cat No N11166 ...
Страница 40: ...Wellwash Versa Main Parts Shaker 38 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 130: ...Maintenance Disposal of the instrument 128 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 142: ...System Log 140 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 144: ...Certificate of Decontamination 142 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 146: ......
Страница 147: ...N11166 Thermo Fisher Scientific Oy Ratastie 2 P O Box 100 FI 01621 Vantaa Finland www thermoscientific com ...