Menus, Tabs and Parameters
Maintenance menu
Thermo Fisher Scientific
Thermo Scientific Wellwash Versa User Manual
101
11.
Press the
F
1
key to continue calibration.
OR
Press the
F
2
key to cancel calibration.
OR
Press the
F
3
key to repeat calibration when the software informs
that the Delta% or accuracy are not acceptable. (The accuracy
must be 2.5–3.5 x that of 100 μl.)
12.
Calibration is now complete. Press the
F
1
key to accept the new
calibration.
Note
Calibration for the wash head is saved and it becomes the active
calibration.
▲
13.
Press the
OK
button when the procedure has finished.
14.
Verify the new performance by dispensing a plate, for example, with
300 μl. Tare weigh the plate, dispense and measure the weight to
determine the actual dispensed volume.
Calculation method: Weight [mg] / 96 = Volume/well [μl]
Содержание Wellwash Versa
Страница 1: ...Thermo Scientific Wellwash Versa User Manual Rev 1 3 ...
Страница 2: ......
Страница 3: ...Thermo Scientific Wellwash Versa User Manual Rev 1 3 Cat No N11166 ...
Страница 40: ...Wellwash Versa Main Parts Shaker 38 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 130: ...Maintenance Disposal of the instrument 128 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 142: ...System Log 140 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 144: ...Certificate of Decontamination 142 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 146: ......
Страница 147: ...N11166 Thermo Fisher Scientific Oy Ratastie 2 P O Box 100 FI 01621 Vantaa Finland www thermoscientific com ...