Menus, Tabs and Parameters
Main menu
Thermo Fisher Scientific
Thermo Scientific Wellwash Versa User Manual
75
●
Aspirate speed
– Define the aspirate rate as Low, Medium, or High.
The default is
High
in 96-well format and
Low
in 384-well format.
●
Aspirate time (s)
– The aspirate time ranges from 0 s to 10 s in
1 s increments. The default is
1 s
.
Caution
A zero second aspirate time will increase the residual
volume.
▲
●
Strip over mode
– Define the strip over function. The default is
No
.
•
No
– All selected strips are processed as fast as possible.
•
Yes
– If
Yes
is selected, unused strips are processed dry. The
time parameters are measured during the first protocol run.
The wash head stays above unused strips for as long as it
takes to process each selected strip. This selection is
recommended when the timing for washing strips is critical.
You can maintain the time between cycles constant with this
option.
A Cell aspirate step removes liquid from the wells. Refer to “Cell wash”
on page 71.
The
Cell aspirate
step has the following parameters:
●
Aspirate height (mm)
– The aspirate height ranges from 0 mm to
14 mm in 0.1 mm increments. The default height varies according
to the plate type and well shape (see Table 6–7 on page 72). Refer to
Figure 6–40.
Caution
Too high an aspirate height increases the residual volume.
▲
Caution
Aspirating too low may disrupt the cells or block the
channels.
▲
●
Aspirate speed
– Define the aspirate rate Low, Medium, or High.
The default is
Low
.
Cell aspirate
Содержание Wellwash Versa
Страница 1: ...Thermo Scientific Wellwash Versa User Manual Rev 1 3 ...
Страница 2: ......
Страница 3: ...Thermo Scientific Wellwash Versa User Manual Rev 1 3 Cat No N11166 ...
Страница 40: ...Wellwash Versa Main Parts Shaker 38 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 130: ...Maintenance Disposal of the instrument 128 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 142: ...System Log 140 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 144: ...Certificate of Decontamination 142 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 146: ......
Страница 147: ...N11166 Thermo Fisher Scientific Oy Ratastie 2 P O Box 100 FI 01621 Vantaa Finland www thermoscientific com ...