Menus, Tabs and Parameters
Maintenance menu
100
Thermo Scientific Wellwash Versa User Manual
Thermo Fisher Scientific
to
8.
Weigh the dispensed plate and enter the mass of the dispensed
liquid in grams using the number keys.
9.
Press the
F
1
key to continue calibration.
OR
Press the
F
2
key to cancel calibration.
OR
Press the
F
3
key to repeat calibration when the software informs
that the Delta% or accuracy are not acceptable. (The accuracy of
the first plate must be ± 50% and that of the second plate must be
± 5%.)
10.
Continue dispensing and weighing the third and fourth plate in
the same way.