Maintenance
Checking the liquid level sensors
118
Thermo Scientific Wellwash Versa User Manual
Thermo Fisher Scientific
Figure 11–44.
Autoclavable and un-autoclavable liquid bottle parts
Warning
When handling the waste bottle, it is advisable to observe
applicable safety precautions, such as the wearing of disposable powder-
free gloves, safety glasses, and protective clothing, to avoid potential
infectious disease contamination.
▲
Check the performance of the liquid level sensors, as described in the
“Liquid bottles and channels” on page 22 and “Performing the
operational check after switching on” on page 31.
You may autoclave, that is, sterilize materials with pressurized steam.
Autoclavable
and
un-autoclavable
parts of the washer unit are listed in
Table 11–9. Refer also to Figure 11–44.
The autoclaving conditions are as follows: at 121°C with 1 bar
underpressure for 20 minutes.
Table 11–9.
Autoclavable and un-autoclavable parts of the washer
Autoclavable
Un-autoclavable
2x8 cell wash head
Aerosol cover
Liquid bottles
Blank wash head
Liquid bottle caps
Liquid filter
Intake tubes
Liquid level sensor assemblies
Priming vessel
Sensor nut
Tube connectors
1x8 wash heads
1x12 wash heads
1x16 wash head
Checking the liquid
level sensors
Autoclavation
Cannot be
autoclaved
Can be
autoclaved
Liquid filter
Sensor cable
Intake tube
Nut
Содержание Wellwash Versa
Страница 1: ...Thermo Scientific Wellwash Versa User Manual Rev 1 3 ...
Страница 2: ......
Страница 3: ...Thermo Scientific Wellwash Versa User Manual Rev 1 3 Cat No N11166 ...
Страница 40: ...Wellwash Versa Main Parts Shaker 38 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 130: ...Maintenance Disposal of the instrument 128 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 142: ...System Log 140 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 144: ...Certificate of Decontamination 142 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 146: ......
Страница 147: ...N11166 Thermo Fisher Scientific Oy Ratastie 2 P O Box 100 FI 01621 Vantaa Finland www thermoscientific com ...