Menus, Tabs and Parameters
Maintenance menu
Thermo Fisher Scientific
Thermo Scientific Wellwash Versa User Manual
93
4.
Press the
F
2
key to abort the soaking and to continue the
procedure.
The priming vessel is emptied.
5.
Press the
OK
button when the procedure has finished.
The procedure moves the wash head down into the priming vessel to the
soak position. You need to add deionized distilled water or a suitable
cleaning agent manually into the priming vessel.
This procedure can be used to leave the wash head in liquid (standby) or
when a clogged wash head is cleaned and the instrument pump cannot
be used to fill the priming vessel.
1.
Select the
Move wash head
row using the
Down
arrow key and
press the
OK
button.