Menus, Tabs and Parameters
Settings menu
Thermo Fisher Scientific
Thermo Scientific Wellwash Versa User Manual
83
•
Volume (ml)
– The autoprime volume ranges from 1 ml to
30 ml in 1 ml increments. The default is
5 ml
.
•
Time (hh:mm)
– The time ranges from 0:00 (= Off) to 9:00
in increments of 15 min. The default is
Off
.
Liquid level sensors are built into the caps of all bottles to avoid
overflow of the waste bottle and to warn the user when the buffer
bottles are almost empty.
The liquid level sensors can be enabled or disabled in the
Sensors
window.
Caution
If the sensors are disabled, there will be no warning prior to the
buffer bottle being empty or the waste bottle full. Failure may result in a
weakened or faulty wash result, liquid spillage, or biohazard.
▲
Select the
Sensors
row in the
Settings
menu and press the
OK
button.
The
Sensors
window has the following parameters:
●
Buffer A sensor
– Disable or enable the buffer A sensor. The default
is
Enabled
.
●
Buffer B sensor
– Disable or enable the buffer B sensor. The default
is
Enabled
.
●
Buffer Rinse sensor
– Disable or enable the buffer Rinse sensor.
The default is
Enabled
.
●
Buffer Waste sensor
– Disable or enable the buffer Waste sensor.
The default is
Enabled
.
●
Plate sensor
– Disable or enable the Plate sensor. The default is
Enabled
. When enabled, the instrument checks whether there is a
plate on the plate carrier at the beginning of the protocol run.
Caution
If the plate sensor is disabled, there will not be a warning
indicating a missing plate. A missing plate may result in spillage.
▲
Sensors
Содержание Wellwash Versa
Страница 1: ...Thermo Scientific Wellwash Versa User Manual Rev 1 3 ...
Страница 2: ......
Страница 3: ...Thermo Scientific Wellwash Versa User Manual Rev 1 3 Cat No N11166 ...
Страница 40: ...Wellwash Versa Main Parts Shaker 38 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 130: ...Maintenance Disposal of the instrument 128 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 142: ...System Log 140 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 144: ...Certificate of Decontamination 142 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 146: ......
Страница 147: ...N11166 Thermo Fisher Scientific Oy Ratastie 2 P O Box 100 FI 01621 Vantaa Finland www thermoscientific com ...