Troubleshooting Guide
USB memory device
134
Thermo Scientific Wellwash Versa User Manual
Thermo Fisher Scientific
Code Explanation
Suggested action
27
Parameter memory write failure.
Contact authorized technical service.
50
File open error. File not found.
–
53
The file already exists.
Use a different file name.
54
The file does not exist.
–
55
The media is full.
Delete unused protocols.
56
End of file.
–
57
Other file error (none of the previous).
–
58
Firmware update aborted by user.
–
59
File read failed.
–
60
File not found.
–
61
USB memory stick not present.
Insert a USB memory device.
62
File write failed.
–
70
Not enough memory.
Delete unused protocols.
71
Adding the step not successful.
–
80
Invalid step parameter(s).
–
81
Too many step parameters.
–
82
At least one of the step parameters is missing.
–
83
Step list is not empty.
–
84
Creating a new step list failed.
–
85
No strips are selected.
Select strips.
86
Invalid SOF parameter.
–
91
USB device timeout.
Try with another USB memory device.
Table 13–13.
Warning codes reported
Code Explanation
Suggested action
101
WAI timer already timed out.
–
102
Liquid level alert.
Check the liquid levels in the bottles.
It is recommended to format the USB memory device if export or
import fails.
It is recommended to use FAT16 or FAT32 formatted USB memory
devices.
It is not recommended to use multiple drive USB memory devices that
contain virtual CD drives.
USB memory
device
Cont.
Содержание Wellwash Versa
Страница 1: ...Thermo Scientific Wellwash Versa User Manual Rev 1 3 ...
Страница 2: ......
Страница 3: ...Thermo Scientific Wellwash Versa User Manual Rev 1 3 Cat No N11166 ...
Страница 40: ...Wellwash Versa Main Parts Shaker 38 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 130: ...Maintenance Disposal of the instrument 128 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 142: ...System Log 140 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 144: ...Certificate of Decontamination 142 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 146: ......
Страница 147: ...N11166 Thermo Fisher Scientific Oy Ratastie 2 P O Box 100 FI 01621 Vantaa Finland www thermoscientific com ...