Running Protocols
Strip selection
48
Thermo Scientific Wellwash Versa User Manual
Thermo Fisher Scientific
Instrument to wash head
12ml
Bottle to instrument
8ml
Bottle to wash head
20ml
Figure 5–34.
Dead volumes of the liquid tubing
Caution
Prime the liquid system with the liquid to be used (wash
buffer). Ensure that the dispensing pump is not run for longer than a
few minutes without liquid, otherwise it may be damaged. Check for
possible air leaks.
▲
Note
If the instrument has not been used for a longer time, you might
need to perform the Boost prime option. Refer to “Service” on page
102.
▲
Note
In case of two-buffer protocols, both the buffer channels have to
be primed before the protocol is started. If channels are switched within
the same protocol, add a Prime step to the protocol to ensure that liquid
in the tubing is replaced completely (a volume of at least 15 ml is
required).
▲
Strip selection can be carried out in two ways:
1.
To select the columns/strips of the plate, use the number keys.
OR
2.
Press the
OK
button on the
Layout
row in the
Main
menu.
Strip selection
Содержание Wellwash Versa
Страница 1: ...Thermo Scientific Wellwash Versa User Manual Rev 1 3 ...
Страница 2: ......
Страница 3: ...Thermo Scientific Wellwash Versa User Manual Rev 1 3 Cat No N11166 ...
Страница 40: ...Wellwash Versa Main Parts Shaker 38 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 130: ...Maintenance Disposal of the instrument 128 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 142: ...System Log 140 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 144: ...Certificate of Decontamination 142 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 146: ......
Страница 147: ...N11166 Thermo Fisher Scientific Oy Ratastie 2 P O Box 100 FI 01621 Vantaa Finland www thermoscientific com ...