Wellwash Versa Main Parts
Shaker
Thermo Fisher Scientific
Thermo Scientific Wellwash Versa User Manual
37
Figure 3–29.
Configurations of Wellwash Versa wash heads
It is possible to shake the microplate during the soak time to improve
washing efficiency. The linear shaker operates at three different speeds
(Table 3–1).
Table 3–1.
Shaking speeds
Speed designation
Speed
Low
5 Hz, amplitude 2.5 mm
Medium
10 Hz, amplitude 1.5 mm
High
15 Hz, amplitude 1 mm
Caution
The instrument is not intended for shaking purposes only. A
separate microplate shaker is available if needed (e.g. Thermo Scientific™
iEMS™ Incubator/Shaker). Refer to