Menus, Tabs and Parameters
Main menu
68
Thermo Scientific Wellwash Versa User Manual
Thermo Fisher Scientific
The wash parameters are grouped into two windows. Use the
Right
and
Left
arrow keys to move from one window to another.
The
Wash
step has the following parameters:
●
Buffer source
– You can define the buffer source channel A, B, or
Rinse. The default is
A
.
●
Wash volume (μl)
– Define the volume of the wash buffer used in
microplate washing. For a 96-well plate, the range is 50 μl to
1000 μl in 10 μl increments. The default value is
300 μl
.
For a 384-well plate, the range is 20 μl to 300 μl in 10 μl
increments. The default value is
100 μl
.
Note
It is possible to dispense more than the well volume and the excess
liquid is aspirated, as the aspiration is always on. If a so-called overflow
volume is selected where the dispensing volume is greater than the well
volume, pay attention to the dispense height. Check that the wash head
tips are inside the well.
▲
●
Wash cycles
– Define the number of wash cycles that will be
performed, from 1 to 10. The default value is
3
.
●
Soak time (mm:ss)
– Define the time period between wash cycles in
the Wash step. The default value is
00:00
(= no soak). The
minimum value is 0 s and the maximum value 60 min in increments
of 1 s.
Содержание Wellwash Versa
Страница 1: ...Thermo Scientific Wellwash Versa User Manual Rev 1 3 ...
Страница 2: ......
Страница 3: ...Thermo Scientific Wellwash Versa User Manual Rev 1 3 Cat No N11166 ...
Страница 40: ...Wellwash Versa Main Parts Shaker 38 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 130: ...Maintenance Disposal of the instrument 128 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 142: ...System Log 140 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 144: ...Certificate of Decontamination 142 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 146: ......
Страница 147: ...N11166 Thermo Fisher Scientific Oy Ratastie 2 P O Box 100 FI 01621 Vantaa Finland www thermoscientific com ...