Operating the Instrument
Main menu
42
Thermo Scientific Wellwash Versa User Manual
Thermo Fisher Scientific
The info text bar also shows the liquid level in the bottles (
A
,
B
,
R
, and
W
). The liquid bottles have sensors to enable liquid level detection. Fill
the buffer bottles and empty the waste bottle if the red bottle icon
appears on the info text bar. Refer to “Sensors” on page 83.
Liquid level detection (LLD):
Liquids A, B, and Rinse
•
icon is
green
when the bottle is filled with liquid
•
icon is
red
with an exclamation mark when the
bottle is nearly empty or disconnected (The warning
is activated when there is approximately 290 ml in
the 2 liter wash bottle / 580 ml in the 4 litter wash
bottle.)
•
icon is
empty
with a cross when the LLD is disabled
Waste
•
icon is
red
with an exclamation mark when the
bottle is full or disconnected
•
icon is
green
when the bottle is not yet full
•
icon is
empty
with a cross when the LLD is disabled
Liquid level
detection
Содержание Wellwash Versa
Страница 1: ...Thermo Scientific Wellwash Versa User Manual Rev 1 3 ...
Страница 2: ......
Страница 3: ...Thermo Scientific Wellwash Versa User Manual Rev 1 3 Cat No N11166 ...
Страница 40: ...Wellwash Versa Main Parts Shaker 38 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 130: ...Maintenance Disposal of the instrument 128 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 142: ...System Log 140 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 144: ...Certificate of Decontamination 142 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 146: ......
Страница 147: ...N11166 Thermo Fisher Scientific Oy Ratastie 2 P O Box 100 FI 01621 Vantaa Finland www thermoscientific com ...