Installation
Operating precautions and limitations before operation
Thermo Fisher Scientific
Thermo Scientific Wellwash Versa User Manual
29
This section describes defects and abnormal stresses.
Warning
If the instrument is not functioning properly, it may create
electromagnetic interference, which could impair the operation of other
devices or equipment in the usual laboratory environment.
▲
Whenever it is likely that the protection system has been impaired, the
instrument should be made inoperative and be secured against any
unintended operation. Contact authorized technical service
immediately.
The protection is likely to be impaired if, for example, the instrument:
●
Shows any visible damage
●
Fails to perform the intended functions
●
Has been subjected to prolonged storage under unfavorable
conditions
●
Has been subjected to severe transport stresses.
1.
Read this manual in its entirety, as it contains information necessary
to ensure safe operation.
2.
Always ensure that the electrical supply in the laboratory conforms
to that specified on the type label at the rear of the instrument.
3.
Ensure that the bottle and wash head tubing are properly fitted.
Ensure that the intake tube reaches completely down to the bottom
of the liquid bottle so that no red tubing is visible. Check regularly
that the liquid filter in the intake tube is attached and clean.
4.
Fill the wash and rinse bottles only after the installation and
operational check of the system.
5.
Check that the liquids are put in the correct bottles called for by the
protocol.
6.
Check that there is sufficient liquid in the liquid bottles and room
in the waste bottle to run the protocol or series of protocols. Empty
the waste bottle before a run or a series of runs.
The liquid level sensor will warn you if safe levels have been passed.
Caution
To ensure proper function, do not disable or disconnect
sensors.
▲
Defects and
abnormal stresses
Operating
precautions and
limitations
before operation
Содержание Wellwash Versa
Страница 1: ...Thermo Scientific Wellwash Versa User Manual Rev 1 3 ...
Страница 2: ......
Страница 3: ...Thermo Scientific Wellwash Versa User Manual Rev 1 3 Cat No N11166 ...
Страница 40: ...Wellwash Versa Main Parts Shaker 38 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 130: ...Maintenance Disposal of the instrument 128 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 142: ...System Log 140 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 144: ...Certificate of Decontamination 142 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 146: ......
Страница 147: ...N11166 Thermo Fisher Scientific Oy Ratastie 2 P O Box 100 FI 01621 Vantaa Finland www thermoscientific com ...