Maintenance
Refitting the transport lock
124
Thermo Scientific Wellwash Versa User Manual
Thermo Fisher Scientific
7.
Place the instrument in a large plastic bag.
8.
Place a cloth soaked in the prepared decontaminant solution into
the bag. Ensure that the cloth does not come into contact with the
instrument.
9.
Seal the bag firmly and leave the instrument in the bag for at least
24 hours.
10.
Remove the instrument from the bag.
11.
After decontamination, clean the instrument using a mild detergent.
12.
After performing this decontamination procedure, enclose a signed
and dated “
Certificate of Decontamination
” (see Appendix B) both
inside the transport package and attached to the outside of the
package.
When you relocate the instrument or ship it for service, make sure that
you refit the transport lock.
1.
Switch off the instrument.
2.
Align the wash head arm and the plate carrier in order to refit the
transport lock screw (Figure 11–47).
Figure 11–47.
Aligning the plate carrier and wash head arm
3.
Screw on the transport lock screw and the transport lock tag with
the supplied Allen key (Figure 11–48).
Refitting the
transport lock
Содержание Wellwash Versa
Страница 1: ...Thermo Scientific Wellwash Versa User Manual Rev 1 3 ...
Страница 2: ......
Страница 3: ...Thermo Scientific Wellwash Versa User Manual Rev 1 3 Cat No N11166 ...
Страница 40: ...Wellwash Versa Main Parts Shaker 38 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 130: ...Maintenance Disposal of the instrument 128 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 142: ...System Log 140 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 144: ...Certificate of Decontamination 142 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 146: ......
Страница 147: ...N11166 Thermo Fisher Scientific Oy Ratastie 2 P O Box 100 FI 01621 Vantaa Finland www thermoscientific com ...