Installation
Setups
18
Thermo Scientific Wellwash Versa User Manual
Thermo Fisher Scientific
Figure 2–8.
Priming vessel installation
Warning
The priming vessel may be contaminated after the instrument
has been used.
▲
The wash heads are packed separately and must be installed before use.
Keep the wash head package for storage purposes. If the wash head is
not used, place it first into the enclosed plastic bag and then into the
original package. The wash heads supplied with the instrument are
calibrated at the factory. The wash head box includes an instrument
serial number according to which it has been calibrated at the factory.
Optional wash heads should be calibrated by the user prior to use.
Warning
The wash heads may be contaminated after the instrument
has been used.
▲
Warning
Failing to connect the dispensing or aspiration tube may cause
spillage.
▲
Caution
Only use wash heads that have an identification label.
▲
Note
Always wear disposable powder-free gloves when handling the
wash heads.
▲
Note
Do not remove the lot number sticker on the wash head.
▲
1.
The instrument is supplied with a 1x8, 2x8, 1x12, or 2x12 wash
head.
2.
Fit the supplied wash heads to the tubing according to Figure 2–9
through Figure 2–11. Note the difference in sizes.
●
The larger aspiration (waste) tube is connected to the
bottom hole.
Installing the wash
head(s)
Содержание Wellwash Versa
Страница 1: ...Thermo Scientific Wellwash Versa User Manual Rev 1 3 ...
Страница 2: ......
Страница 3: ...Thermo Scientific Wellwash Versa User Manual Rev 1 3 Cat No N11166 ...
Страница 40: ...Wellwash Versa Main Parts Shaker 38 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 130: ...Maintenance Disposal of the instrument 128 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 142: ...System Log 140 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 144: ...Certificate of Decontamination 142 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 146: ......
Страница 147: ...N11166 Thermo Fisher Scientific Oy Ratastie 2 P O Box 100 FI 01621 Vantaa Finland www thermoscientific com ...