Running Protocols
Starting a run
Thermo Fisher Scientific
Thermo Scientific Wellwash Versa User Manual
51
To start a run:
1.
Press the
START
button to start the run.
The microplate will be processed according to the protocol.
Note
Ensure that the liquid channels work properly during dispensing
and aspiration.
▲
2.
Remove the plate after the protocol has ended.
Note
Prime the instrument with deionized distilled water at the end of
the batch.
▲
To cancel a run:
Press the
STOP
button to cancel a run.
Note
It is not possible to stop the run when a user input is required as a
response to a warning or error message.
▲
The
RINSE
button starts a sequence where the rinse liquid is pumped
through the liquid system for a short period of time and left in the
tubing. The rinse solution is normally deionized distilled water.
The volume used during rinsing is defined in the Settings > Prime &
rinse parameters >
Rinse volume (Rinse button) (ml)
menu. Refer to
“Prime & rinse parameters” on page 81.
Starting a run
Canceling a run
Rinsing the
liquid system
Содержание Wellwash Versa
Страница 1: ...Thermo Scientific Wellwash Versa User Manual Rev 1 3 ...
Страница 2: ......
Страница 3: ...Thermo Scientific Wellwash Versa User Manual Rev 1 3 Cat No N11166 ...
Страница 40: ...Wellwash Versa Main Parts Shaker 38 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 130: ...Maintenance Disposal of the instrument 128 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 142: ...System Log 140 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 144: ...Certificate of Decontamination 142 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 146: ......
Страница 147: ...N11166 Thermo Fisher Scientific Oy Ratastie 2 P O Box 100 FI 01621 Vantaa Finland www thermoscientific com ...