Installation
Setups
26
Thermo Scientific Wellwash Versa User Manual
Thermo Fisher Scientific
Figure 2–20.
Liquid bottles connected to the instrument
Warning
Make sure that the liquid level of the waste bottle is always
kept below the maximum level indicated on the bottle to avoid potential
overflow, as the contents of the waste bottle is potentially infectious.
▲
Warning
The contents of the waste bottle is potentially infectious, so it
is important to wear protective clothing, such as disposable gloves, a
laboratory coat, and safety glasses, when emptying or handling the waste
bottle.
▲
Caution
Always ensure that the liquid bottles are attached properly.
Always connect the bottle cap to the correct bottle and to the correct
connector on the left side of the instrument. Otherwise the wash
performance can be seriously affected.
▲
Caution
Ensure that the intake tube reaches fully down to the bottom
of the liquid bottle so that no red tubing is visible.
▲
Caution
Check regularly that the liquid filters in the intake tubes are
clean. If not, replace or clean them.
▲
Caution
To ensure proper function, do not disconnect the sensor cables
from the instrument during operation.
▲
Содержание Wellwash Versa
Страница 1: ...Thermo Scientific Wellwash Versa User Manual Rev 1 3 ...
Страница 2: ......
Страница 3: ...Thermo Scientific Wellwash Versa User Manual Rev 1 3 Cat No N11166 ...
Страница 40: ...Wellwash Versa Main Parts Shaker 38 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 130: ...Maintenance Disposal of the instrument 128 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 142: ...System Log 140 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 144: ...Certificate of Decontamination 142 Thermo Scientific Wellwash Versa User Manual Thermo Fisher Scientific ...
Страница 146: ......
Страница 147: ...N11166 Thermo Fisher Scientific Oy Ratastie 2 P O Box 100 FI 01621 Vantaa Finland www thermoscientific com ...