Operation
Instrument Controls Menu
3-48
Model 48
i
Instruction Manual
Thermo Fisher Scientific
EXT ALARM 3
O2 CONC MAX
O2 CONC MIN
The Non-Alarm status screen allows the user to select the non-alarm status
for the selected relay output. The selected item is shown by “<--” after it.
●
In the Main Menu, choose Instrument Controls > I/O Configuration >
Output Relay Settings > select Relay > Instrument State >
Non-Alarm
.
NON ALARM STATUS ITEMS:
>NONE
AUTORANGE
LOCAL/REMOTE
SERVICE
UNITS
RANGE AVG DIAGS ALARM
ZERO MODE
SPAN MODE
SAMPLE MODE
PURGE MODE
The Digital Input Settings menu displays a list of the 16 digital inputs
available, and allows the user to select the logic state and instrument
parameter for the relay selected.
Note
The digital inputs must be asserted for at least one second for the
action to be activated.
▲
Note
Not all of the I/O available in the instrument is brought out on the
supplied terminal board. If more I/O is desired, an alternative means of
connection is required. (See optional
Terminal Block and Cable Kits
.)
▲
●
In the Main Menu, choose Instrument Controls > I/O Configuration >
Digital Input Setting
s
.
DIGITAL INPUT SETTINGS:
>1 NOP SET BACKGROUND
2 NOP CAL TO LO SPAN
3 NOP AOUTS TO ZERO
4 NOP AOUTS TO FS
5 NOP NONE
RANGE AVG DIAGS ALARM
Non-Alarm
Digital Input Settings
Содержание 48i
Страница 1: ...Model 48i Instruction Manual Gas Filter Correlation CO Analyzer Part Number 101891 00 10Aug2017 ...
Страница 4: ......
Страница 6: ......
Страница 37: ...Installation Connecting External Devices Thermo Fisher Scientific Model 48i Instruction Manual 2 9 ...
Страница 144: ......
Страница 154: ......
Страница 216: ......
Страница 228: ......
Страница 237: ...Optional Equipment Mounting Options Thermo Fisher Scientific Model 48i Instruction Manual 9 9 Figure 9 4 Bench Mounting ...
Страница 240: ......
Страница 321: ......