C-Link Protocol Commands
Communications Configuration
B-44
Model 48
i
Instruction Manual
Thermo Fisher Scientific
Table B–9.
Reply Termination Formats
Format
Reply Termination
00 <CR>
01
<NL> sum xxxx <CR>
where xxxx = 4 hexadecimal digits that represent the sum of all the
characters (bytes) in the message
host name
This command reports the host name string. The following example
reports the host name is set to
i
Series. This command returns “bad cmd” if
no host name has been set.
Send:
host
name
Receive: host
name
iSeries
set host name
string
This command sets the host name
string
, where
string
is 1-13 alphanumeric
characters (with no spaces). The following example sets the host name to
analyzer01. This command returns “bad cmd” if no host name has been
set.
Send:
set host name analyzer01
Receive:
set host name analyzer01 ok
instr name
This command reports the instrument name. The following example
reports the instrument name for the Model 48
i
.
Send:
instr
name
Receive: instr
name
CO
Analyzer
CO
Analyzer
instrument id
This command reports the instrument identification (ID). The following
example reports the current setting of the instrument ID.
Send:
instrument
id
Receive:
instrument id 48
set instrument id value
This command sets the instrument id to
value
, where
value
is an integer
between 0 and 127 inclusive. The following example shows the instrument
ID changed to 50.
Содержание 48i
Страница 1: ...Model 48i Instruction Manual Gas Filter Correlation CO Analyzer Part Number 101891 00 10Aug2017 ...
Страница 4: ......
Страница 6: ......
Страница 37: ...Installation Connecting External Devices Thermo Fisher Scientific Model 48i Instruction Manual 2 9 ...
Страница 144: ......
Страница 154: ......
Страница 216: ......
Страница 228: ......
Страница 237: ...Optional Equipment Mounting Options Thermo Fisher Scientific Model 48i Instruction Manual 9 9 Figure 9 4 Bench Mounting ...
Страница 240: ......
Страница 321: ......