Optional Equipment
Internal Oxygen (O2) Sensor
9-2
Model 48
i
Instruction Manual
Thermo Fisher Scientific
Figure 9–1.
Flow Diagram, Internal O
2
Sensor
This section describes how to calibrate the optional internal O
2
sensor.
Be sure to power the analyzer for half an hour at normal operating
conditions (with the cover on) before calibration. The internal O
2
sensor
should be calibrated at the zero and span points, as described in the
following procedure.
Note
The total airflow of the calibration gases must exceed the total
demand of the analyzer. The Model 48
i
requires approximately 1000
cc/min of sample flow, so a total airflow of at least 1100 cc/min is
recommended.
▲
Use the following procedure to calibrate the oxygen sensor under normal
conditions. If you are unable to calibrate the oxygen sensor effectively using
the following procedure, use the Alternate Calibration Procedure.
Calibration Equipment:
Test gas: Nitrogen zero gas - span gas approximately 20% oxygen
concentration.
1.
Supply the nitrogen zero gas to the SAMPLE port on the rear of the
analyzer and wait for the O
2
reading to stabilize.
Internal O
2
Sensor
Calibration
Содержание 48i
Страница 1: ...Model 48i Instruction Manual Gas Filter Correlation CO Analyzer Part Number 101891 00 10Aug2017 ...
Страница 4: ......
Страница 6: ......
Страница 37: ...Installation Connecting External Devices Thermo Fisher Scientific Model 48i Instruction Manual 2 9 ...
Страница 144: ......
Страница 154: ......
Страница 216: ......
Страница 228: ......
Страница 237: ...Optional Equipment Mounting Options Thermo Fisher Scientific Model 48i Instruction Manual 9 9 Figure 9 4 Bench Mounting ...
Страница 240: ......
Страница 321: ......