Servicing
I/O Expansion Board (Optional) Replacement
Thermo Fisher Scientific
Model 48
i
Instruction Manual
7-35
2.
Unplug the I/O expansion board cable from the EXPANSION BD
connector on the motherboard.
3.
Remove the two standoffs holding the I/O expansion board connector
to the rear panel (
Figure 7–16
).
4.
Pop off the board from the mounting studs and remove the board.
5.
To install the I/O expansion board, follow the previous steps in reverse.
Figure 7–15.
Replacing the I/O Expansion Board (Optional)
Figure 7–16.
Rear Panel Board Connectors
Motherboard
Digital Output Board
I/O Expansion Board (Opt)
Motherboard Support Bracket
Mounting Studs
Motherboard
Standoffs (2)
I/O Expansion Board
Digital Output Board