Operation
Instrument Controls Menu
Thermo Fisher Scientific
Model 48
i
Instruction Manual
3-25
The Calibrate O
2
Background screen is used to adjust the background, or
perform a “zero calibration.” Be sure the analyzer samples nitrogen until
the readings stabilize. The display shows the current O
2
reading.
●
In the Main Menu, choose Calibration >
Calibrate O
2
Background
.
O2 BACKGROUND:
O2: 0.12 %
CURRENTLY: 0.00 % ?
SET O2 TO ZERO
RANGE AVG DIAGS ALARM
The Calibrate O
2
Coefficient screen is used to adjust the O
2
span
concentration while sampling span gas of known concentration.
The display shows the current O
2
concentration reading. The next line of
the display is where the O
2
calibration gas concentration is entered.
●
In the Main Menu, choose Calibration >
Calibrate O
2
Coefficient
.
CALIBRATE O2:
O2: 20.80 %
SPAN CONC:
00020.8
0
MOVE CURSOR
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
The Instrument Controls menu contains a number of items that may be
selected to control various instrument operational parameters. The
firmware controls listed in this menu enable control of the listed
instrument functions. The O
2
Compensation menu item is only displayed
if the internal O
2
sensor option is installed.
●
In the Main Menu, choose
Instrument Controls
.
Calibrate O
2
Background
Calibrate O
2
Coefficient
Instrument Controls
Menu
Содержание 48i
Страница 1: ...Model 48i Instruction Manual Gas Filter Correlation CO Analyzer Part Number 101891 00 10Aug2017 ...
Страница 4: ......
Страница 6: ......
Страница 37: ...Installation Connecting External Devices Thermo Fisher Scientific Model 48i Instruction Manual 2 9 ...
Страница 144: ......
Страница 154: ......
Страница 216: ......
Страница 228: ......
Страница 237: ...Optional Equipment Mounting Options Thermo Fisher Scientific Model 48i Instruction Manual 9 9 Figure 9 4 Bench Mounting ...
Страница 240: ......
Страница 321: ......