Operation
Instrument Controls Menu
3-46
Model 48
i
Instruction Manual
Thermo Fisher Scientific
●
In the Main Menu, choose Instrument Controls >
I/O Configuration
.
I/O CONFIGURATION:
>OUTPUT RELAY SETTINGS
DIGITAL INPUT SETTINGS
ANALOG OUTPUT CONFIG
ANALOG INPUT CONFIG
RANGE AVG DIAGS ALARM
The Output Relay Settings menu displays a list of the 10 digital output
relays available, and allows the user to select the logic state or instrument
parameter for the relay selected.
Note
The digital outputs may take up to one second after the assigned
state occurs to show up on the outputs.
▲
●
In the Main Menu, choose Instrument Controls > I/O Configuration >
Output Relay Settings
.
OUTPUT RELAY SETTINGS:
>1 NOP GEN ALARM
2 NOP NONE
3 NOP UNITS
4 NOP CONC ALARM
5 NOP NONE
RANGE AVG DIAGS ALARM
The Logic State menu item is used to change the selected I/O relay to
either normally open or normally closed. The default state is open, which
indicates that a relay connected between the digital output pin and ground
is normally open and closes to trigger the digital output action.
●
Press
to toggle and set the logic state open or closed.
OUTPUT RELAY SETUP:
>LOGIC STATE OPEN
INSTRUMENT STATE
RANGE AVG DIAGS ALARM
Output Relay Settings
Logic State
Содержание 48i
Страница 1: ...Model 48i Instruction Manual Gas Filter Correlation CO Analyzer Part Number 101891 00 10Aug2017 ...
Страница 4: ......
Страница 6: ......
Страница 37: ...Installation Connecting External Devices Thermo Fisher Scientific Model 48i Instruction Manual 2 9 ...
Страница 144: ......
Страница 154: ......
Страница 216: ......
Страница 228: ......
Страница 237: ...Optional Equipment Mounting Options Thermo Fisher Scientific Model 48i Instruction Manual 9 9 Figure 9 4 Bench Mounting ...
Страница 240: ......
Страница 321: ......