Servicing
Analog Output Testing
Thermo Fisher Scientific
Model 48
i
Instruction Manual
7-29
The Set Analog Outputs screen appears.
4.
Press
to set the output to zero.
The Output Set To line displays Zero.
5.
Check that the meter is displaying the zero value. If the meter reading
differs by more than one percent of the full-scale output, the analog
outputs should be adjusted. Refer to the “
Analog Output Calibration
”
procedure that follows.
6.
Press
to set the output to full-scale.
The Output Set To line displays Full-Scale.
7.
Check that the meter is displaying a full-scale value. If the meter
reading differs by more than one percent of the full-scale output, the
analog outputs should be adjusted. Refer to the “
Analog Output
Calibration
” procedure that follows.
8.
Press
to reset the analog outputs to normal.
Figure 7–14.
Rear Panel Analog Input and Output Pins
Current Outputs
Analog Voltage Inputs
Analog Voltage Outputs