MODBUS Protocol
MODBUS Parameters Supported
Thermo Fisher Scientific
Model 48
i
Instruction Manual
C-11
Register Number
Variable
49&50
NOT USED
51&52
NOT USED
53&54
NOT USED
55&56
BIAS SUPPLY
57&58
NOT USED
59&60
NOT USED
61&62
NOT USED
63&64
NOT USED
65&66
NOT USED
67&68
EXT ALARMS
69&70
O2 %*
71&72
O2 SENS TEMP*
73&74
NOT USED
75&76
NOT USED
77&78
NOT USED
79&80
CO COR*
81&82
LO CO COR*
83&84
HI CO COR*
85&86
NOT USED
87&88
NOT USED
89&90
NOT USED
*If internal O2 Sensor is installed
IMPORTANT NOTE
The addresses in the following tables are Protocol
Data Unit (PDU) addresses. Verify the coil number on your MODBUS
master to ensure that it matches the coil number on the instrument.
▲
Note
Writing 1 to the coil number shown in the following table will
initiate the “action triggered” listed in the table. This state must be held for
at least 1 second to ensure the instrument detects the change and triggers
the appropriate action.
▲
Note
The coils within each coil group in the following table are mutually
exclusive and will not be triggered if there is conflict. Before you assert (1)
one coil in a group, make sure the other coils in the group are de-asserted
(0).
▲
Содержание 48i
Страница 1: ...Model 48i Instruction Manual Gas Filter Correlation CO Analyzer Part Number 101891 00 10Aug2017 ...
Страница 4: ......
Страница 6: ......
Страница 37: ...Installation Connecting External Devices Thermo Fisher Scientific Model 48i Instruction Manual 2 9 ...
Страница 144: ......
Страница 154: ......
Страница 216: ......
Страница 228: ......
Страница 237: ...Optional Equipment Mounting Options Thermo Fisher Scientific Model 48i Instruction Manual 9 9 Figure 9 4 Bench Mounting ...
Страница 240: ......
Страница 321: ......