Operation
Service Menu
3-82
Model 48
i
Instruction Manual
Thermo Fisher Scientific
Note
An independent flow sensor is required to read the flow, then the
operator enters the flow value on this screen to perform the calibration.
▲
Note
Wait at least 30 seconds for the reading to stabilize before saving the
value.
▲
●
In the Main Menu, choose Service > Flow Calibration >
Span
.
CALIBRATE FLOW SPAN:
CURRENTLY: 0.800 L/min
SET TO: 0.75
0
L/min ?
MOVE CURSOR
CHANGE VALUE SAVE
RANGE AVG DIAGS ALARM
The Restore Default Flow Calibration screen allows the user to reset the
flow calibration configuration values to factory defaults.
●
In the Main Menu, choose Service > Flow Calibration >
Set Defaults
.
The Initial Sample/Reference (S/R) Ratio screen displays both the initial
S/R ratio and the current S/R ratio. The initial S/R ratio is determined at
the factory, and is used to correct for the slight variations found from one
correlation wheel to another. The only time the initial S/R ratio should be
changed is when the correlation wheel is replaced or sample/reference is
between 1.14 and 1.18.
Note
This adjustment should only be performed by an instrument service
technician.
▲
●
In the Main Menu, choose Service >
Initial S/R Ratio
.
Restore Default Flow
Calibration
Initial
Sample/Reference
Ratio
RESTORE DEFAULT CAL:
RESTORE
RANGE AVG DIAGS ALARM
RESTORE DEFAULT CAL:
RESTORE
ARE YOU SURE YOU WANT TO?
x
PRESS
TO CONFIRM RESTORE
RANGE AVG DIAGS ALARM
Содержание 48i
Страница 1: ...Model 48i Instruction Manual Gas Filter Correlation CO Analyzer Part Number 101891 00 10Aug2017 ...
Страница 4: ......
Страница 6: ......
Страница 37: ...Installation Connecting External Devices Thermo Fisher Scientific Model 48i Instruction Manual 2 9 ...
Страница 144: ......
Страница 154: ......
Страница 216: ......
Страница 228: ......
Страница 237: ...Optional Equipment Mounting Options Thermo Fisher Scientific Model 48i Instruction Manual 9 9 Figure 9 4 Bench Mounting ...
Страница 240: ......
Страница 321: ......