Preventive Maintenance
Pump Rebuilding
Thermo Fisher Scientific
Model 48
i
Instruction Manual
5-5
In order to check for leaks across the optional valves, plug the SPAN inlet
line, press
until span appears in the status line and follow the
“
External Leaks
” procedure.
If the pressure drops below 250 mmHg the valve associated with the span
line is functioning normally. Repeat for the valve associated with the zero
line by plugging the zero inlet, press
until zero appears in the status
line and follow the “
External Leaks
” procedure.
If the pressure drops below 250 mmHg, the valve associated with the zero
line is functioning normally.
For new technology pump, use the following procedure to rebuild the
pump (
Figure 5–2
). The new technology pump will have the repair kit part
number on its label. If no repair kit part number is listed, refer to the
procedure for the older model pump following this procedure. To replace
the pump, see “
Pump Replacement
” on page
7-20
.
Equipment Required:
Flatblade screwdriver
Pump rebuild kit (flapper valve and diaphragm)
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component.
▲
1.
Turn instrument OFF, unplug the power cord, and remove the cover.
Make a sketch of the position of any tubes and fittings for ease of re-
assembly later.
2.
Mark the position of the pump top plate, bottom plate and compressor
housing relative to each other by drawing a line on the edges with a
pencil or marker to ensure proper re-assembly.
3.
The aluminum compressor housing cover must be removed to gain
access to the inside compressor housing. Remove the four screws and
then remove the cover. Re-use any gasketing. Remove any debris that
may have accumulated in the bottom of the compressor housing.
Leaks Across the Optional
Zero/Span and Sample
Solenoid Valves
Pump Rebuilding
Disassembly
Содержание 48i
Страница 1: ...Model 48i Instruction Manual Gas Filter Correlation CO Analyzer Part Number 101891 00 10Aug2017 ...
Страница 4: ......
Страница 6: ......
Страница 37: ...Installation Connecting External Devices Thermo Fisher Scientific Model 48i Instruction Manual 2 9 ...
Страница 144: ......
Страница 154: ......
Страница 216: ......
Страница 228: ......
Страница 237: ...Optional Equipment Mounting Options Thermo Fisher Scientific Model 48i Instruction Manual 9 9 Figure 9 4 Bench Mounting ...
Страница 240: ......
Страница 321: ......