Operation
Calibration Menu
Thermo Fisher Scientific
Model 48
i
Instruction Manual
3-23
The Zero/Span Period Hours screen defines the period or interval between
zero/span checks. Periods between 0 and 999 hours are acceptable. To turn
the zero/span check off, set the period to 0.
●
In the Main Menu, choose Calibration > Zero/Span Check >
Period
Hr
.
ZERO/SPAN PERIOD:
CURRENTLY: 024 HRS
SET TO: 02
5
HRS ?
MOVE CURSOR
CHANGE VALUE SAVE
RANGE AVG DIAGS ALARM
The Total Duration Hour line is the sum of the zero, span, and purge
duration minutes, and is for viewing only.
The Zero Duration Minutes screen defines how long zero air is sampled by
the instrument. The Span and Purge Duration Minutes screens look and
function the same way as the zero duration screen. The span duration
screen is used to set how long the span gas is sampled by the instrument.
The purge duration screen is used to set how long the purge period will be
after doing a zero or span check. This gives the instrument time to flush
out the zero and span gas before any meaningful data is taken. Logged data
is flagged as taken during a purge to show that the data is suspect.
Durations between 0 and 99 minutes are acceptable. Each time a zero/span
check occurs the zero check is done first, followed by the span check. To
perform just a zero check, set the span duration to 0 (off). The same applies
to perform just a span check.
●
In the Main Menu, choose Calibration > Zero/Span Check >
Zero,
Span
or
Purge Duration Min
.
ZERO DURATION:
CURRENTLY: 30 MIN
SET TO: 31 MIN ?
MOVE CURSOR
CHANGE VALUE SAVE
RANGE AVG DIAGS ALARM
Period Hours
Total Duration Hour
Zero/Span/Purge
Duration Minutes
Содержание 48i
Страница 1: ...Model 48i Instruction Manual Gas Filter Correlation CO Analyzer Part Number 101891 00 10Aug2017 ...
Страница 4: ......
Страница 6: ......
Страница 37: ...Installation Connecting External Devices Thermo Fisher Scientific Model 48i Instruction Manual 2 9 ...
Страница 144: ......
Страница 154: ......
Страница 216: ......
Страница 228: ......
Страница 237: ...Optional Equipment Mounting Options Thermo Fisher Scientific Model 48i Instruction Manual 9 9 Figure 9 4 Bench Mounting ...
Страница 240: ......
Страница 321: ......